Abstract
PURPOSE: To provide a tool to enable gamma analysis software algorithms to be included in a quality assurance (QA) program.
METHODS: Four image sets were created comprising two geometric images to independently test the distance to agreement (DTA) and dose difference (DD) elements of the gamma algorithm, a clinical step and shoot IMRT field and a clinical VMAT arc. The images were analysed using global and local gamma analysis with 2 in-house and 8 commercially available software encompassing 15 software versions. The effect of image resolution on gamma pass rates was also investigated.
RESULTS: All but one software accurately calculated the gamma passing rate for the geometric images. Variation in global gamma passing rates of 1% at 3%/3mm and over 2% at 1%/1mm was measured between software and software versions with analysis of appropriately sampled images.
CONCLUSION: This study provides a suite of test images and the gamma pass rates achieved for a selection of commercially available software. This image suite will enable validation of gamma analysis software within a QA program and provide a frame of reference by which to compare results reported in the literature from various manufacturers and software versions.
Original language | English |
---|---|
Pages (from-to) | 568-73 |
Number of pages | 6 |
Journal | Radiotherapy and Oncology |
Volume | 118 |
Issue number | 3 |
Early online date | 26 Feb 2016 |
DOIs | |
Publication status | Published - Mar 2016 |