Abstract
PURPOSE: To provide a tool to enable gamma analysis software algorithms to be included in a quality assurance (QA) program.
METHODS: Four image sets were created comprising two geometric images to independently test the distance to agreement (DTA) and dose difference (DD) elements of the gamma algorithm, a clinical step and shoot IMRT field and a clinical VMAT arc. The images were analysed using global and local gamma analysis with 2 in-house and 8 commercially available software encompassing 15 software versions. The effect of image resolution on gamma pass rates was also investigated.
RESULTS: All but one software accurately calculated the gamma passing rate for the geometric images. Variation in global gamma passing rates of 1% at 3%/3mm and over 2% at 1%/1mm was measured between software and software versions with analysis of appropriately sampled images.
CONCLUSION: This study provides a suite of test images and the gamma pass rates achieved for a selection of commercially available software. This image suite will enable validation of gamma analysis software within a QA program and provide a frame of reference by which to compare results reported in the literature from various manufacturers and software versions.
| Original language | English |
|---|---|
| Pages (from-to) | 568-73 |
| Number of pages | 6 |
| Journal | Radiotherapy and Oncology |
| Volume | 118 |
| Issue number | 3 |
| Early online date | 26 Feb 2016 |
| DOIs | |
| Publication status | Published - Mar 2016 |
Fingerprint
Dive into the research topics of 'A tool to include gamma analysis software into a quality assurance program'. Together they form a unique fingerprint.Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver