Abstract
Objective
To report the diagnostic test accuracy of dipstick urinalysis for the detection of urinary tract infections (UTIs) in febrile infants aged 90 days or less attending the emergency department (ED).
Design
Retrospective cohort study.
Patients
Febrile infants aged 90 days or less attending between 31 August 2018 and 1 September 2019.
Main outcome measures
The sensitivity, specificity and predictive values of dipstick urinalysis in detecting UTIs defined as growth of ≥100 000 cfu/mL of a single organism and the presence of pyuria (>5 white blood cells per high-power field).
Setting
Eight paediatric EDs in the UK/Ireland.
Results
A total of 275 were included in the final analysis. There were 252 (92%) clean-catch urine samples and 23 (8%) were transurethral bladder catheter samples. The median age was 51 days (IQR 35–68.5, range 1–90), and there were 151/275 male participants (54.9%). In total, 38 (13.8%) participants had a confirmed UTI. The most sensitive individual dipstick test for UTI was the presence of leucocytes. Including ‘trace’ as positive resulted in a sensitivity of 0.87 (95% CI 0.69 to 0.94) and a specificity of 0.73 (95% CI 0.67 to 0.79). The most specific individual dipstick test for UTI was the presence of nitrites. Including trace as positive resulted in a specificity of 0.91 (95% CI 0.86 to 0.94) and a sensitivity of 0.42 (95% CI 0.26 to 0.59).
Conclusion
Point-of-care urinalysis is moderately sensitive and highly specific for diagnosing UTI in febrile infants. The optimum cut-point to for excluding UTI was leucocytes (1+), and the optimum cut-point for confirming UTI was nitrites (trace).
To report the diagnostic test accuracy of dipstick urinalysis for the detection of urinary tract infections (UTIs) in febrile infants aged 90 days or less attending the emergency department (ED).
Design
Retrospective cohort study.
Patients
Febrile infants aged 90 days or less attending between 31 August 2018 and 1 September 2019.
Main outcome measures
The sensitivity, specificity and predictive values of dipstick urinalysis in detecting UTIs defined as growth of ≥100 000 cfu/mL of a single organism and the presence of pyuria (>5 white blood cells per high-power field).
Setting
Eight paediatric EDs in the UK/Ireland.
Results
A total of 275 were included in the final analysis. There were 252 (92%) clean-catch urine samples and 23 (8%) were transurethral bladder catheter samples. The median age was 51 days (IQR 35–68.5, range 1–90), and there were 151/275 male participants (54.9%). In total, 38 (13.8%) participants had a confirmed UTI. The most sensitive individual dipstick test for UTI was the presence of leucocytes. Including ‘trace’ as positive resulted in a sensitivity of 0.87 (95% CI 0.69 to 0.94) and a specificity of 0.73 (95% CI 0.67 to 0.79). The most specific individual dipstick test for UTI was the presence of nitrites. Including trace as positive resulted in a specificity of 0.91 (95% CI 0.86 to 0.94) and a sensitivity of 0.42 (95% CI 0.26 to 0.59).
Conclusion
Point-of-care urinalysis is moderately sensitive and highly specific for diagnosing UTI in febrile infants. The optimum cut-point to for excluding UTI was leucocytes (1+), and the optimum cut-point for confirming UTI was nitrites (trace).
Original language | English |
---|---|
Number of pages | 5 |
Journal | Archives of Disease in Childhood |
Early online date | 24 Aug 2022 |
DOIs | |
Publication status | Early online date - 24 Aug 2022 |