Preparation of a radon-free thoron source for a thoron calibration chamber

Ferenc Fabian, Anita Csordas, Zoltan Sas, Tibor Kovacs

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)
123 Downloads (Pure)

Abstract

During long-term radon/thoron surveys, solid-state nuclear track detectors (SSNTDs) are commonly used. For the application of these integral devices, it is important to solve the commensurate calibration. At the calibration process, radon can be interfering with thoron; therefore, radon-free thoron atmosphere is required for the accurate calibration. In this study, radon-free thoron source with a high emanation capability was prepared by chemical separation of 228Ac. Based on the results of this and the previously performed experiments, a method combines precipitation with FeCl3 and embedding into ceramic matrix containing low feldspar content and heat treatment on 200 °C has been proposed.
Original languageEnglish
Number of pages7
JournalJournal of Radioanalytical and Nuclear Chemistry
DOIs
Publication statusEarly online date - 20 Jul 2016
Externally publishedYes

Keywords

  • thoron source
  • calibration
  • Ac-228 separation