The cemented custom femoral stem--a 10 year review

  • S O'Brien
  • , R K Wilson
  • , B M Hanratty
  • , N W Thompson
  • , M E Wallace
  • , J R Nixon
  • , D W Engela
  • , J F Orr
  • , G H Isaac
  • , D E Beverland

Research output: Contribution to journalArticlepeer-review

Abstract

We report a series of 706 patients (759 hip implants) with an average follow up of 10.5 years (range, 10-11 years) following total hip replacement (THR) using a cemented custom-made femoral stem and a cemented HDP acetabular component. The fate of every implant is known. One hundred and seventy-four patients (23%) were deceased at the time of their 10-year review all died with a functioning THR in situ. Four hundred and sixty-two patients (61%) were subsequently reviewed. One hundred and twenty three patients (16%) were assessed by telephone review, as they were too ill or unwilling to attend. Kaplan-Meier survival analysis (all components) demonstrated a median survival at 10 years of 96.05% or 95% Confidence Intervals (CI) for median survival of (94.41% to 97.22%). Revision surgery occurred in 30 cases (3.9%). Seventeen had full revisions (2.2%) and 13 (1.7%) socket revisions only. Twenty-one out of 30 revisions were for infection or dislocation. There were 2 cases (0.3%) of revision for aseptic loosening of the stem. The 10-year results of the custom femoral titanium stem are encouraging and compare well with other cemented systems.
Original languageEnglish
Pages (from-to)194-204
Number of pages11
JournalEmirates Journal for Engineering Research
Volume17
Issue number4
Publication statusPublished - 2007

Fingerprint

Dive into the research topics of 'The cemented custom femoral stem--a 10 year review'. Together they form a unique fingerprint.

Cite this