The quality of reporting of diagnostic accuracy studies in glaucoma using the Heidelberg retina tomograph

M. Shunmugam, A. Azuara-Blanco

Research output: Contribution to journalArticlepeer-review

17 Citations (Scopus)

Abstract

PURPOSE. Scanning laser tomography with the Heidelberg retina tomograph (HRT; Heidelberg Engineering, Heidelberg, Germany) has been proposed as a useful diagnostic test for glaucoma. This study was conducted to evaluate the quality of reporting of published studies using the HRT for diagnosing glaucoma. METHODS. A validated Medline and hand search of English-language articles reporting on measures of diagnostic accuracy of the HRT for glaucoma was performed. Two reviewers selected and appraised the papers independently. The Standards for Reporting of Diagnostic Accuracy (STARD) checklist was used to evaluate the quality of each publication. RESULTS. A total of 29 articles were included. Interobserver rating agreement was observed in 83% of items (? = 0.76). The number of STARD items properly reported ranged from 5 to 18. Less than a third of studies (7/29) explicitly reported more than half of the STARD items. Descriptions of key aspects of the methodology were frequently missing. For example, the design of the study (prospective or retrospective) was reported in 6 of 29 studies, and details of participant sampling (e.g., consecutive or random selection) were described in 5 of 29 publications. The commonest description of diagnostic accuracy was sensitivity and specificity (25/29) followed by area under the ROC curve (13/29), with 9 of 29 publications reporting both. CONCLUSIONS. The quality of reporting of diagnostic accuracy tests for glaucoma with HRT is suboptimal. The STARD initiative may be a useful tool for appraising the strengths and weaknesses of diagnostic accuracy studies.
Original languageEnglish
Pages (from-to)2317-2323
Number of pages7
JournalInvestigative ophthalmology & visual science
Volume47
Issue number6
DOIs
Publication statusPublished - 01 Jun 2006

Fingerprint Dive into the research topics of 'The quality of reporting of diagnostic accuracy studies in glaucoma using the Heidelberg retina tomograph'. Together they form a unique fingerprint.

Cite this