The validation of Fibit ZipTM physical activity monitor as a measure of free-living physical activity

Mark A Tully, Cairmeal McBride, Leonnie Heron, Ruth F Hunter

Research output: Contribution to journalArticlepeer-review

136 Citations (Scopus)
314 Downloads (Pure)

Abstract

BACKGROUND: The new generation of activity monitors allow users to upload their data to the internet and review progress. The aim of this study is to validate the Fitbit Zip as a measure of free-living physical activity.

FINDINGS: Participants wore a Fitbit Zip, ActiGraph GT3X accelerometer and a Yamax CW700 pedometer for seven days. Participants were asked their opinion on the utility of the Fitbit Zip. Validity was assessed by comparing the output using Spearman's rank correlation coefficients, Wilcoxon signed rank tests and Bland-Altman plots. 59.5% (25/47) of the cohort were female. There was a high correlation in steps/day between the Fitbit Zip and the two reference devices (r = 0.91, p < 0.001). No statistically significant difference between the Fitbit and Yamax steps/day was observed (Median (IQR) 7477 (3597) vs 6774 (3851); p = 0.11). The Fitbit measured significantly more steps/day than the Actigraph (7477 (3597) vs 6774 (3851); p < 0.001). Bland-Altman plots revealed no systematic differences between the devices.

CONCLUSIONS: Given the high level of correlation and no apparent systematic biases in the Bland Altman plots, the use of Fitbit Zip as a measure of physical activity. However the Fitbit Zip recorded a significantly higher number of steps per day than the Actigraph.

Original languageEnglish
JournalBMC Research Notes
Volume7
Issue number952
DOIs
Publication statusPublished - 23 Dec 2014

Fingerprint

Dive into the research topics of 'The validation of Fibit ZipTM physical activity monitor as a measure of free-living physical activity'. Together they form a unique fingerprint.

Cite this